4-21
   
1. Remove vertical column. Refer to Section 4-1.
2. Remove retaining ring from bottom of column base shaft. See Illustration 4-12.
ILLUSTRATION 4-12
   
COLUMN BEARING PLATE
COLUMN BASE SHAFT
LOWER COLUMN BEARING
UPPER COLUMN BEARING
3. Pull lower column bearing from column base shaft.
4. Remove screws and washers securing lower brake to column bearing plate.
5. Remove lower brake from column base shaft.