6-2
2. Verify that AMX key switch is OFF, circuit breaker is OFF, and charger cord is
unplugged.
3. Remove battery ground cable marked #2 from the ground stud. Tape to ensure
it does not short. SEPARATE IT FROM ALL OTHER CABLES.
 All battery lead connectors must be insulated to prevent electrical contact
with the frame or other battery terminals.
4. Tape cable marked #5 separately from ALL other cables.
5. Remove all remaining cables from ground stud. See Illustration 6-1. Tape each
cable separately. Keep ALL cables separate.
6. Remove battery lead cable marked #4 from terminal on Circuit Breaker in
AMX1 A3 Module. Be certain boot covers the connector and it cannot short out.
Tape connector to ensure it will not short.
7. Remove battery compartment cover capscrews and cover.
8. Remove insulator board from compartment entrance.
      
1. If unit has a battery sense harness, disconnect harness from J11 of the battery test
terminal strip. See Illustration 6-2.
2. Disconnect sense leads from J9 and J10 on left hand side of the battery test termiĆ
nal strip.

Table of Contents

Other manuals for GE AMX 4+

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE AMX 4+ and is the answer not in the manual?

GE AMX 4+ Specifications

General IconGeneral
TypeMobile X-ray System
Maximum kVp125 kVp
CollimatorManual Collimator
BatteryRechargeable, lead-acid
Image Receptor CompatibilityCR and DR systems
Focal Spot Size1.2 mm
mAs Range0.04 - 320 mAs
X-ray TubeRotating Anode

Related product manuals