4-14
 Connect Brake
1. Insert brake cable through top of channel on back of column and down into terĆ
minal box.
2. Connect upper brake cable leads to terminal block. See Illustration 4-2.
3. Replace terminal box cover located on back of vertical column and secure with
two screws.
4. Install main and safety cables on the carriage cable mounting strut. Refer to SecĆ
tion 4-4.
5. Install vertical column. Refer to Section 4-2.
 Remove Upper Pulley, Bearings and Brake
(Alternate)
 This procedure can be used to replace the brake without removing the vertical
column
1. Remove horizontal arm. Refer to Section 3-5.
D It is not necessary to remove yoke from arm.
2. Disconnect upper brake. Refer to Section 4Ć4Ć1.
3. Disconnect main and safety cables from carriage. Refer to Section 4Ć6.
4. Move carriage to bottom of column.
5. Remove four capscrews securing pulley and brake assembly to top of column.
6. Remove the Heyco bushing in column where brake cable passes through column
and enters pulley and brake assembly.
D Bushing extends into pulley and brake casting.
7. Remove pulley and brake assembly from top of column while sliding the brake
cable through slot in top of column.
D This must be done together or the brake cable may be damaged.
8. Remove screws holding upper brake to the mounting bracket.

Table of Contents

Other manuals for GE AMX 4+

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE AMX 4+ and is the answer not in the manual?

GE AMX 4+ Specifications

General IconGeneral
TypeMobile X-ray System
Maximum kVp125 kVp
CollimatorManual Collimator
BatteryRechargeable, lead-acid
Image Receptor CompatibilityCR and DR systems
Focal Spot Size1.2 mm
mAs Range0.04 - 320 mAs
X-ray TubeRotating Anode

Related product manuals