EasyManuals Logo

GE AMX 4+ Technical Manual

GE AMX 4+
132 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #26 background imageLoading...
Page #26 background image
 
        
  
1-8
ILLUSTRATION 1-6
  
 
15. Examine the last exposure and determine which direction to move the field lamp
to best position it around the x-ray field.
16. Turn on the field lamp.
 "%&()& # &(#''
$"%$##(' & $( # (  #&* ! !"% )'("#( & $ #$(
!$$ &(!* ( ( ! !"% (' &(#'' ' !##
17. The field lamp adjustment area will be hot so use a metal tool to move the bracket
when repositioning the light field area on the film cassette.
18. When the light field area has been repositioned equidistant around the x-ray
field, tighten the adjustment screws. Alternately tightening the two screws will
help prevent the bracket from slipping out of place.
19. Repeat steps 1 through 16 until the percentage of SID" requirements are obĆ
tained.
$( If the percentage of SID requirement cannot be obtained, the light field is
either too large or in rare instances too small. In this case the collimator will
have to be returned to the factory.
20. Replace the back cover and fasten it in place with four metric Phillips head
screws. See Illustration 1-4.
21. After any adjustment to the field light area, the collimator crosshair window posiĆ
tion should be checked and corrected if necessary. Refer to Section 1-5.


Table of Contents

Other manuals for GE AMX 4+

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE AMX 4+ and is the answer not in the manual?

GE AMX 4+ Specifications

General IconGeneral
TypeMobile X-ray System
Maximum kVp125 kVp
CollimatorManual Collimator
BatteryRechargeable, lead-acid
Image Receptor CompatibilityCR and DR systems
Focal Spot Size1.2 mm
mAs Range0.04 - 320 mAs
X-ray TubeRotating Anode

Related product manuals