3-10
7. Return arm to upright position.
8. Slide the center and inner tubes from the outer tube.
9. Place center and inner tubes in upside down position.
10. Remove four screws holding latch block to center tube. Remove block. This proĆ
vides access to the mounting plate holding screw. See Illustration 3-6.
11. Remove mounting plate capscrew holding brake and cable assembly to the center
tube.
12. Remove the two brake capscrews holding the brake & cable assembly to the rear
of the center tube.
13. Remove two or three screws (depends on unit you have) holding the connector
bracket and ground wire to the center tube.
14. Remove two retaining rings from inner shaft on center tube. Slide shaft out perĆ
mitting removal of two spacers and two bearings.
15. Carefully slide the brake and cable assembly, and the inner tube from the center
tube.
16. Remove two capscrews holding the brake and cable assembly bracket to the inner
tube.
17. Remove brake and cable assembly from inner tube.

Table of Contents

Other manuals for GE AMX 4+

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE AMX 4+ and is the answer not in the manual?

GE AMX 4+ Specifications

General IconGeneral
TypeMobile X-ray System
Maximum kVp125 kVp
CollimatorManual Collimator
BatteryRechargeable, lead-acid
Image Receptor CompatibilityCR and DR systems
Focal Spot Size1.2 mm
mAs Range0.04 - 320 mAs
X-ray TubeRotating Anode

Related product manuals