GE DRAFT VOLUSON™ P8/VOLUSON™P6
DIRECTION 5775469, R
EVISION 3 DRAFT (JULY 19, 2018) BASIC SERVICE MANUAL
Chapter 7 - Diagnostics/Troubleshooting 7-55
NOTE: BIOS value needs to be changed again after simulation mode for normal scan mode as following
procedure.
1.) Turn the system on and press “Del” key to enter the BIOS menu.
2.) Enter the password “Polaris” and press “Enter” key.
3.) Press “F9” and select “Yes”.
4.) Press “F10” and select “Yes” to save the change.
Operation:
1.) Power On / Boot Up the system as described in Section 3-6-2 "Power On / Boot Up" on page 3-25.
2.) As soon as the start-up process bar indicates "Initialization of Voluson", press the H key on the
Alphanumeric keyboard.
3.) Keep the H key pressed during complete startup (~ 1 minute). Following message appears:
4.) Click on CONTINUE. The "Probe Selection" menu appears on the screen.
5.) Press the FREEZE
key. The application starts, but does not show any Ultrasound information.
6.) Backup all images (see:Section 4-5-6 "Archiving Images" on page 4-22).
Figure 7-48 Start-up screen
Figure 7-49 H Key pressed during startup