GE RAFT VOLUSON™ P8/VOLUSON™P6
DIRECTION 5775469, R
EVISION 3 DRAFT (JULY 19, 2018) BASIC SERVICE MANUAL
7-56 Section 7-8 - Minimum Configuration to Boot/Scan
7.) Save a "Full Backup" as described in Section 4-5-3 "Save Full System Configuration (Full Backup)"
on page 4-17) or press the ALT+D
key to gather detailed information and data from the system.
(see: Section 7-4-3-1 "Export System Data (by pressing the ALT + D key)" on page 7-16).
8.) Press the F2
key on the control panel and select SYSTEM SETUP to invoke setup desktop.
9.) Select ADMINISTRATION
, open the SYSTEM INGO page and click REBOOT MACHINE.
Figure 7-50 Reboot Machine