EasyManua.ls Logo

GE Voluson P6 - Page 371

GE Voluson P6
620 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE DRAFT VOLUSON™ P8/VOLUSON™P6
DIRECTION 5775469, R
EVISION 3 DRAFT (JULY 19, 2018) BASIC SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-9
10.)Verify that only the DVD drive is connected to the system, then click OK (Figure 8-4 below).
11.)To start update procedure click YES
The system saves Full Backup in silent mode on R:, then it reboots into LINUX.
A silent “Rollback” image from C:\ is stored on R:\.
After executing all LINUX commands, the system reboots again.
NOTE: If the currently installed software has to be upgraded to a newer version, a new software specific
“Permanent key is required.
12.)Enter the appropriate “Permanent Key”
(calculated in OKOS), select OK and confirm with ENTER.
If the entered Permanent key is correct, following window appears, confirm with ENTER
.
Figure 8-4 Verify that USB devices are disconnected, then click OK
Figure 8-5 Inserting Permanent Key
Figure 8-6 Confirming Permanent Key
Note: This Example shows
fictional numbers.

Table of Contents

Other manuals for GE Voluson P6

Related product manuals