RemediesPossible causesAlarm Message
Check patient and breathing system.
Check mask/ prongs size and patient fit.
Leakage too high. The mask / prongs
may not be adjusted properly for the
patient or may be the wrong size.
Leakage out of
range (A high
priority alarm when
NIV disconnect flow
is enabled). See
section Medium
priority alarms.
If possible, connect to AC power supply.
Replace and discard all batteries if this
message appears even when batteries
are fully charged.
Battery voltage too low. Cannot
guarantee continued ventilator operation.
Low battery voltage
Connect to AC power.Less than 3 minutes left of battery
operation.
No battery capacity
Check O
2
cell / sensor and connection.O
2
cell / sensor missing or disconnected.O
2
cell / sensor
failure
Check air supply.
Perform a Pre-use check.
Perform O
2
cell adaptation.
Measured O
2
concentration exceeds the
set value by more than 5 Vol.%.
Gas supply or air line disconnected.
No supply from wall outlet.
The air gas module is disconnected.
If no gas is available, then both expiratory
and safety valves will open.
O
2
concentration:
High
Check O
2
supply line.
Perform a Pre-use check.
Perform O
2
cell adaptation.
Measured O
2
concentration is below the
set value by more than 5 Vol.% or
concentration is below 18 Vol.% which
is independent of operator settings.
Gas delivered in O
2
supply line is not O
2
.
O
2
sensor faulty or exhausted.
O
2
cell uncalibrated.
O
2
/oxygen gas module faulty.
O
2
concentration:
Low
Check patient and breathing system.
Check ventilator settings and alarm
limits.
Airway pressure exceeds preset Upper
Pressure Limit.
Kinked or blocked tubing.
Mucus or secretion plug in endotracheal
tube or in airways.
Patient coughing or fighting ventilator.
Inspiratory flow rate too high.
Improper alarm setting.
Blocked expiratory filter.
Paw high
Restart the ventilator and perform a
Pre-use check.
Contact a service technician.
Software error.Restart ventilator!
Contact a service technician.During Pre-use check the system found
problems with the opening pressure for
the safety valve.
Safety valve test
failed
136
SERVO-s VENTILATOR SYSTEM V8.0, User's Manual
| 9
| System messages |

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Getinge Maquet SERVO-s and is the answer not in the manual?

Getinge Maquet SERVO-s Specifications

General IconGeneral
BrandGetinge
ModelMaquet SERVO-s
CategoryFan
LanguageEnglish

Related product manuals