RemediesPossible causesAlarm Message
Restart the ventilator and perform a
Pre-use check.
Check ventilator settings.
Software error, memory corrupt.Settings lost;
Restart ventilator
Perform a Pre-use check.
Change the expiratory cassette and
perform a Pre-use check.
Contact a service technician.
Technical problem with the expiratory
cassette.
Technical error in
Expiratory cassette
Restart the ventilator, perform a Pre-use
check and check all settings.
Contact a service technician.
Ventilator settings lost.Technical error:
Restart ventilator
Check patient and breathing system.Time in waiting position is exceeded.
Patient is not connected to the ventilator
or leakage is excessive.
Time in waiting
position exceeds
2 min.
SERVO-s VENTILATOR SYSTEM V8.0, User's Manual
137