WR1000-UM-251019380 6 21
6. INSPECTION AND CHECK PROCEDURES
Fig. 6-2-26. Load Settings menu
(4) Specify the parameters on the Load Settings menu to the same settings that you specified in Step (2),
then read out the saved measurement conditions. Check that the measurement conditions that you
changed in Step (3) have returned to their previous settings.
6.2.11 Chart Feeding Precision
Preparation
Prepare a glass scale that is at least 30 cm long.
Setting Procedure
(1) Press the main control panels
MODE
key to select RECORDER as the measurement mode.
(2) Press the
RECORD
key on the Conditions panel to respectively display the Record Settings and Format
Settings windows. At each window, use the arrow keys to move the cursor to the desired parameter, then
use the
ENTER
key and arrow keys to enter the following settings.
Fig. 6-2-27. Record Settings window