Model 5340A
Manual Changes and Options
b. In
Table 6
-
1, change parts
lists
for A10
as
follows:
Change series
1252A to 1220A.
Change A10C4 to “0180
-
0291 C:FXD ELECT 1.0 UF 10% 35VDCW, Mfr Code 56289,
Mfr part number
150D105X9035A2-DYS.”
Add “AlOCR3 1901
-
0040 DI0DE:SILICON 50MA 30WV, Mfr code 07263, Mfr part
number
FDG1088.”
Change AlOCR4 and AlOCR5 to “0122-0057 C:VAR 20PF/.4V, Mfr code 28480, Mfr
part number
0122-0057.”
Add “Note: A1OCR4, AlOCR5, A12CR4, and A12CR5 are
a matched set. If any one diode
is
replaced all four diodes must be replaced.”
Change
A10L4 to “9100-2266 COIL/INDUCTOR FXD 8.2UF lo%, Mfr code 82142,
Mfr part number 09
-
1316
-
2K.”
Change
A10R2 to “0698-3443 RFXD MET FLM 287 OHM
1%
1/8W, Mfr code 28480,
Mfr part number
0698-3443.”
Change A10R5 to “0698-3158 RFXD MET FLM 23.7K OHM
1%
1/8W, Mfr code 28480,
Mfr
part
number 0698-3158.”
Change A10R7 to “0698-3151 R.FXD MET FLM 2.87K OHM
1%
1/8W, Mfr code
28480, Mfr
part
number 0698-3151.”
Change A10RE
to
“0698-0084 R.FXD MET FLM 2.15K OHM
1%
1/8W, Mfr code
28480, Mfr part number
0698-0084.”
Change A10R9 to “0698-3157 RFXD MET FLM 19.6K OHM
1%
1/8W, Mfr code
28480, Mfr part number
0698-3157.”
Add “AlOR10 0757
-
0443 RFXD MET FLM ll.OK OHM
1%
1/8W, Mfr code 28480,
Mfr part number
0757-0443.”
Add “AlOR11 0757
-
0439 R:FXD MET FLM 6.81K OHM
1%
1/8W, Mfr code 28480,
Mfr part number
0757-0439.”
Change A10R24 to “0757-0283 R:FXD MET FLM 2.00K OHM
1%
1/8W, Mfr code
28480, Mfr part number
0757-0283.”
CHANGE
12
a.
On Figure
8
-
34
and 8
-
35, change
F1
to “2.5 amp slow
-
blow 115V operation.” For instru
-
ments with serial prefix 1244A, 2.5 amp fuses were installed, for replacement see Table 6
-
1.
In Table 6
-
1, change
F1
to “2110
-
0015 FUSE:CARTRIDGE 2.5 AMP 125 V
MAX
SLOW
-
BLOW
.”
b.
Change part number of MP6
to
“05340
-
00002.”
Change part number of MP7 to “05340
-
20028.”
c.
On Figure 8
-
22, delete pin W and connecting wire to XA19B.
This
wire was added
to
accommodate future changes.
On Figure 2
-
2 and Figure 3
-
4, change item
1
(10
MHz OUTPUT) to “TIME BASE OUT-
PUT.” Change item
3
(10 MHz INPUT) to “OSC” only.
d.
I
CHANGE
13
This instrument configurations for series 1244A and series 1236A are identical.
’
1,
I/
.I
I
7
-
5