Step 3. To set up the logic analyzer
1
Press
Autoscale
.
2
Press
Logic levels
.
3
Press the
Set
softkey to select
Chan 0-7
.
4
Press the
User
softkey, then use the Entry knob to set the threshold
voltage to the 50% level of the mark generator pulse, based on the
voltages you measured previously.
5
Set the Time/div knob to 20 ns/div.
Testing, Adjusting, and Troubleshooting the Analyzer
Step 3. To set up the logic analyzer
176