Table 5
Readout Settings
Setting Indicator Meaning Application
Time t1 Time at cursor t1 with respect to trigger. Measure delta t of waveform events with
respect to trigger event, or measure period
or frequency for irregular waveforms.
t2 Time at cursor t2 with respect to trigger.
delta t Time between t1 and t2.
1/delta t Frequency corresponding to the measured
delta t.
Binary Bin (t1) Binary value of all channels in the order
15-0 (where 15 is MSB) at cursor t1, plus
edge information where applicable.
Gather binary state information for a group
of signals at a particular waveform event.
Bin (t2) Binary value of all channels in the order
15-0 (where 15 is MSB) at cursor t2, plus
edge information where applicable.
Hex hex (t1) Hexadecimal value of channels that are on
in the order they are displayed, at cursor t1,
plus edge information where applicable.
The MSB is the channel at the top of the
display. If any edges are found, the
character is displayed as an up arrow,
down arrow, or double arrow for rising,
falling, or both edges respectively.
Gather hexadecimal state information for a
group of signals at a particular waveform
event.
hex (t2) Hexadecimal value of channels that are on
in the order they are displayed, at cursor t2,
plus edge information where applicable.
The MSB is the channel at the top of the
display. If any edges are found, the
character is displayed as an up arrow,
down arrow, or double arrow for rising,
falling, or both edges respectively.
Making Analyzer Measurements
To use the cursors
95