To turn channels on and off
To turn off a specific channel, press
On/Off
, select a channel
using the Select knob, then press the leftmost softkey until
Off
is
highlighted.
To turn on a specific channel, press
On/Off
, select a channel
using the Select knob, then press the leftmost softkey until
On
is
highlighted.
If you are already in the
On/Off
menu, you can press the
On/Off
key to toggle a particular channel on or off. The list of channels that are
off is in the upper-left corner of the display. You can also use the first
softkey to select a channel when the On/Off softkey menu is present.
To turn channels 0 through 7 on, press
On/Off
, then press the
On
softkey under “Chan 0 - 7.”
To turn channels 0 through 7 off, press
On/Off
, then press the
Off
softkey under “Chan 0 - 7.”
A corresponding set of On and Off softkeys is available for channels 8 through
15.
The analyzer does not disable input for channels that are off. It simply does
not display those channels. So, you can assign labels to channels that are off
or use them in a trigger specification. Also, if you turn a channel on after an
acquisition, the data acquired for that channel will be displayed. A channel
must be on, though, if you wish to make measurements using the single- or
dual-channel measurement features.