INFICON Routine Operation | 10
074-594-P1H Micro GC Fusion Operating Manual 271 / 319
CAUTION
Over-tightening the sample line nut may damage the thread on the GC
module.
9
Repeat steps 2 to 8 if replacing both modules.
10
When all the GC modules are replaced, replace the rear foam.
11
Replace the side foam. Ensure the dovetail joint is oriented as shown.