ARB Subsystem
Copyright © ITECH Electronic Co., Ltd. 85
Syntax:
LIST:STEP <NR1>,<string>
Query syntax:
LIST:STEP? <NR1>
Example
Three mode:
LIST:STEP
1,"11,11,11,11,11,11,5,5,5,6,6,6,PHASe,90,90,90,16,16,16,2,2,2,Sine,Sine,Sine,
TIME,2,OFF,10,10,10,2,2,2"
Single mode or reverse mode:
LIST:STEP 1,"1,1,10,5,PHAse,90,0,0,Sine,TIME,1,OFF,10,2"
Note: The values set in this command include all parameters of CC and CR
modes, but when returned, only the value displayed in the interface is returned.
LIST:STEP:ITEM <NR1>,<NR1>,<NRf+>
This command is used to configure and query the phase parameters of a certain
step in the list.
<NR1>: Step index , [1,100]
<NR1>:
IAC_A 0
IAC_B 1
IAC_C 2
IAC_SLEW_A 3
IAC_SLEW_B 4
IAC_SLEW_C 5
IDC_A 6
IDC_B 7
IDC_C 8
IDC_SLEW_A 9
IDC_SLEW_B 10
_SLEW_C 11
PHASE_MODE 12
START_A 13
START_B 14
START_C 15
R_A 16
R_B 17
R_C 18