EasyManua.ls Logo

ITech IT8200 Series

ITech IT8200 Series
137 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SURGesag Subsystem
Copyright © ITECH Electronic Co., Ltd. 92
Query syntax:
SURGesag:PHASe:STARt? [MAXimum|MINimum|DEFault]
Returns:
<NRf+>
Example
SURGesag:PHASe:STARt 90 //Set the initial phase of the surge/trap wave
to 90°.
SURGesag:PHASe:STARt? //Query the presently set starting phase.
SURGesag:PHASe:STARt? MAX // Query the maximum value of the starting
phase.
SURGesag:PHASe:WIDTh <NRf+>
This command is used to set and query the surge/trap wave angle width.
<NRf+>: angle width setting value
Syntax:
SURGesag:PHASe:WIDTh <NRf+>
Arguments
<NRf+>
单位
Degree (°)
*RST
0
Query syntax:
SURGesag:PHASe:WIDTh? [MAXimum| MINimum|DEFault]
Returns:
<NRf+>
Example
SURGesag:PHASe:WIDTh 320.0 //Set the angular width to 320°.
SURGesag:PHASe:WIDTh? //Query the set value of angle width.
SURGesag:PHASe:WIDTh? MIN // Query the minimum value of the angle
width.
SURGesag:ACTion <action>
This command is used to set and query the trigger action of the surge/trap wave.

Table of Contents

Other manuals for ITech IT8200 Series