Query syntax:
SURGesag:PHASe:STARt? [MAXimum|MINimum|DEFault]
Returns:
<NRf+>
Example
SURGesag:PHASe:STARt 90 //Set the initial phase of the surge/trap wave
to 90°.
SURGesag:PHASe:STARt? //Query the presently set starting phase.
SURGesag:PHASe:STARt? MAX // Query the maximum value of the starting
phase.
SURGesag:PHASe:WIDTh <NRf+>
This command is used to set and query the surge/trap wave angle width.
<NRf+>: angle width setting value
Syntax:
SURGesag:PHASe:WIDTh <NRf+>
Arguments
<NRf+>
单位
Degree (°)
*RST
0
Query syntax:
SURGesag:PHASe:WIDTh? [MAXimum| MINimum|DEFault]
Returns:
<NRf+>
Example
SURGesag:PHASe:WIDTh 320.0 //Set the angular width to 320°.
SURGesag:PHASe:WIDTh? //Query the set value of angle width.
SURGesag:PHASe:WIDTh? MIN // Query the minimum value of the angle
width.
SURGesag:ACTion <action>
This command is used to set and query the trigger action of the surge/trap wave.