Load Protect Subsystem
Copyright © ITECH Electronic Co., Ltd. 60
Query syntax:
POWer:PROTection:DELay? [MAXimum|MINimum]
Returns:
<MINimum-MAXimum>
Example
POWer:PROTection:DELay 1
[SOURce:]VOLTage:UNDer:PROTection:STATe
<Boolean>
Sets the state of the undervoltage protection. Setting is only required when in DC
mode.
Syntax:
POWer:PROTection:STATe <Boolean>
Arguments
0|1|OFF|ON
Query syntax:
VOLTage:UNDer:PROTection:STATe?
Returns:
0|1|OFF|ON
Example
VOLTage:UNDer:PROTection:STATe 1
[SOURce:]VOLTage:UNDer:PROTection[:LEVel]
<NRf+>
Sets the undervoltage protection value in the DC mode of the machine. The unit is
V. It needs to be set only when in DC mode.
Syntax:
VOLTage:UNDer:PROTection[:LEVel] <NRf+>
Arguments
<MINimum-MAXimum>
Query syntax:
VOLTage:UNDer:PROTection[:LEVel]? [MAXimum|MINimum]