EasyManua.ls Logo

Johnson Controls SABROE Unisab III Control - Software Calibration, Hydraulic Slide Systems

Johnson Controls SABROE Unisab III Control
322 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
7.5.13 Software calibration, hydraulic slide systems
When the capacity/Vi slide position transmitter has been adjusted (see subsections 7.5.1. to
7.5.12.) or if it is not possible to calibrate the transmitter itself, this function can be used to
adjust/fine-tune the signal. It is possible to software calibrate both capacity and Vi slide posi-
tion in the same picture. Select Setup/Configuration/Calibrate/Position.
Minimum position (0%) should always be calibrated before maximum position (100%).
Move the capacity or Vi slide to minimum position as described in subsections 7.5.2. to 7.5.4.
You can move the capacity slide down in position by pressing the
F3
/CAP. DOWN button. If
the actual reading (cap. slide position or Vi slide position) is not showing 0.0%, but e.g.
1.5%, position the cursor at Cap. Zero Adjustm. or Vi zero adjustment and change the cali-
bration value to -1.5%.
Then move the capacity or Vi slide to maximum position as described in subsections 7.5.2. to
7.5.4. You can move the capacity slide up in position by pressing the
F2
/CAP.UP button. If the
actual reading (cap. slide position or Vi slide position) is not showing 100.0%, but e.g. 95.8%,
position the cursor at Cap.100% Adjustm. or Vi 100% adjustment and change the calibration
value to +4.2%.
If recalibration is needed it is recommended to reset both Zero Adjustm. and 100% Adjustm.
to 0.0 before recalibration is carried out.
Note: If the capacity/Vi transmitters cannot be calibrated the calibration values used in Soft-
ware adjustment are usually higher. Eg the values used for a SAB193L are approximately
-46.8% for Cap. Zero Adjustm. and +49.3% for Cap. 100% Adjustm. Similarly the volume
slide values for a SAB 193L are approximately -37.7% for VI Zero Adjustment and +61.8%
for VI 100% Adjustment.
Calibration
248/319
Engineering manual - Unisab III 1.10
001930 en 2014.09

Table of Contents

Related product manuals