PR-23 instrucon manual
Preventing automatic wash:
The preconditioning and wash relays are never activated by the automatic wash con-
trol:
Under the diagnostic message NO SAMPLE (see Section 8.2.6) as this indicates a
clean prism in an empty process line. The diagnostic message is WASH STOP/NO
SAMPLE.
If a wash stop input switch is closed (see Section 6.3.2), indicating e.g. that there
is no process low. The diagnostic message is EXTERNAL WASH STOP.
If the process temperature limit is activated and the temperature falls below the
limit, indicating that the process is not running. The diagnostic message is LOW
TEMP WASH STOP