EasyManuals Logo

K-Patents PR-23 User Manual

K-Patents PR-23
224 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #65 background imageLoading...
Page #65 background image
6 Conguraon and calibraon
55
Theother
sensor
washing?
NO
YES
MANUAL
WASH been
pressed?
NO
YES
Remote
wash
requested?
1
NO
YES
Wash
interval
reached?
YES
NO
Temp
limit
active?
YES
NO
Temp
above
limit?
YES
NO
Message
NOSAMPLE?
2
NO
YES
Washcycle
Restartwash
intervaltimer
NOTES
1. Remotewashistriggeredat
theclosingoftheswitch.If
theswitchisheldclosed,
onlyonewashcycleis
carriedout.
2. Thewashisinhibitedif
thereisnosample,no
sensororthesensorcannot
measurecorrectly.
Figure 6.15 Wash logic

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the K-Patents PR-23 and is the answer not in the manual?

K-Patents PR-23 Specifications

General IconGeneral
BrandK-Patents
ModelPR-23
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals