Keysight B2900 User’s Guide, Edition 5 4-27
Front Panel Reference
Config key group
Save, Recall Saves/recalls all setup information of the B2900. Five storage areas are available in
the built-in non-volatile memory (NVRAM) for this purpose.
Press the Save or Recall softkey to display the softkeys #1, #2, #3, #4, and #5 used
for selecting the storage area. Press a softkey to save or recall the setup information.
Output Filter dialog box
This dialog box provides the following parameters for setting the output filter.
Ch Only on 2-channel models. Channel 1 (Ch 1) or 2 (Ch 2)
This field specifies the channel set by this dialog box.
Filter State Output filter, ON or OFF
Set the filter ON to obtain clean source output without spikes
and overshooting. Note, however, that using a filter may
increase the SMU settling time.
Automatic Filter Automatic filter, ON or OFF
Set the function ON to automatically set the output filter which
provides the best filter characteristics and cutoff frequency.
Time Constant Filter time constant, 5 s to 500 s

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies B2900 Series and is the answer not in the manual?

Keysight Technologies B2900 Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelB2900 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals