EasyManuals Logo

Keysight Technologies B2900 Series User Manual

Keysight Technologies B2900 Series
234 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #141 background imageLoading...
Page #141 background image
Keysight B2900 User’s Guide, Edition 5 4-41
Front Panel Reference
Result key group
Max. Y-axis maximum value of the graph.
Min. Y-axis minimum value of the graph.
Limit Test Result dialog box
This dialog box provides the following GUI for displaying the limit test result. The
result data is shown in the data list area below the Length field.
Ch Only on 2-channel models. Channel 1 (Ch 1) or 2 (Ch 2)
This field specifies the channel of the data to display.
Length Data length
The limit test data contains the following information.
(aaaaa) BIN: bb DATA: +c.ccccccE+dd
(aaaaa) Data index aaaaa
BIN: Bin number bb (01 to 12)
The bin numbers 1 to 12 are associated with the index numbers
1 to 12. See “Limit Test Setup dialog box” on page 4-34.
If the limit test data is out of the bins, 00 is set for the
GRADING mode and 15 is set for the SORTING mode.
DATA: Limit test data +c.ccccccE+dd