4-30 Keysight B2900 User’s Guide, Edition 5
Front Panel Reference
Config key group
Ranging dialog box
This dialog box provides the following parameters for setting details of the auto
ranging operation for measurements.
Ch Only on 2-channel models. Channel 1 (Ch 1) or 2 (Ch 2)
This field specifies the channel set by this dialog box.
Current Auto Ranging Current measurement auto range operation, NORMAL,
SPEED, or RESOLUTION (RESOLN). See Table 4-6.
Voltage Auto Ranging Voltage measurement auto range operation, NORMAL,
SPEED, or RESOLUTION (RESOLN). See Table 4-6.
Threshold Sets the rate value of the formula shown below.
Table 4-6 Measurement Auto Range Operation
Basic operation
The channel automatically sets the range which provides the best resolution in
performing the measurement.
Upward changing operation
If measured data value1, the range changes up after the measurement.
value1 = measurement range rate / 100
Downward changing operation
If measured data value2, the range changes down immediately.
value2 = measurement range rate / 1000
Description
NORMAL Supports basic operation and downward changing operation
described below
SPEED Supports basic operation and upward and downward changing
operations described below
RESOLN Supports basic operation and upward changing operation
described below

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies B2900 Series and is the answer not in the manual?

Keysight Technologies B2900 Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelB2900 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals