5-24 Keysight B2900 User’s Guide, Edition 5
Front Panel Operations
To Execute Measurement
4. When MANUAL is specified, the aperture time setup field and PLC setup field
are displayed. Enter and fix a value in either of the fields to set the measurement
time.
To Set Measurement Range
For details on this function, see “Ranging Mode” on page 6-4 and “Range
parameters” on page 4-12.
1. Press the View key to display the Single view. If the Range parameters are not
displayed, press the Hide XXXX assist key.
2. Move the field pointer to the Measure Volts (voltage measurement range),
Measure Amps (current measurement range), or Measure Ohms (resistance
measurement range) field.
3. Press the rotary knob to change the pointer status to EDIT.
4. Press AUTO to set the auto range operation or FIXED to set the fixed range
operation. The pointer status is changed to MOVE.
5. Move the field pointer to the range value setup field (on the right side of the
Measure Volts, Measure Amps, or Measure Ohms field).
6. Press the rotary knob to change the pointer status to EDIT.
7. Press an assist key to set the following value.
For the auto range operation: Minimum range of the range operation
For the fixed range operation: Range used for measurement
To Perform Sweep Measurement
For details on this function, see “Sweep Output” on page 6-9.
1. Press the View key to display the Single view.
2. Set the source function (see “To Set Source Output Mode” on page 5-13).
3. Set the Source value and Limit (Compliance) value (see “To Apply DC
Voltage/Current” and “To Set Limit (Compliance) Value” on page 5-13).
4. Set the measurement mode (see “To Set Measurement Mode” on page 5-23).
5. Set the sweep source (see “To Set Sweep Output” on page 5-15).
6. Press the View key to display the Graph view.
7. Press the On/Off switch for channel 1 or 2 (Ch1 or Ch2).

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies B2900 Series and is the answer not in the manual?

Keysight Technologies B2900 Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelB2900 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals