Keysight B2900 User’s Guide, Edition 5 4-31
Front Panel Reference
Config key group
Wait Control dialog box
This dialog box provides the following parameters for setting the source and
measurement wait time.
The source wait time is defined as the time the source channel cannot change the
output after the start of a DC output or the trailing edge of a pulse.
The measurement wait time is defined as the time the measurement channel cannot
start measurement after the start of a DC output or the trailing edge of a pulse.
Ch Only on 2-channel models. Channel 1 (Ch 1) or 2 (Ch 2)
This field specifies the channel set by this dialog box.
State Wait time, ON or OFF
Automatic Automatic wait time, ON or OFF
Gain and Offset Parameters for calculating the wait time. See the following
formula.
If State=ON and Automatic=ON:
wait time = Gain initial wait time + Offset
If State=ON and Automatic=OFF:
wait time = Offset
If State=OFF:
wait time=0
The initial wait time is automatically set by the instrument, and cannot be changed.