1-6 Keysight B2900 User’s Guide, Edition 5
Getting Started
Performing Sweep Measurement
Performing Sweep Measurement
B2900 supports sweep measurements. The following example applies a staircase
sweep voltage and measures the current at each step voltage.
Step 1. Press the key to display the Single view.
Step 2. Set the Source mode, Limit value, and measurement mode as shown in “Applying
DC Output” and “Performing Spot Measurement”.
This example sets the voltage output, a 10 mA limit value, and both current and
voltage measurements.
Step 3. Setting the Sweep parameters
1. Press the
More assist key to change the assist keys.
2. Press the
Show Sweep assist key to display the Sweep setup parameters.
3. Press the rotary knob to change the status to EDIT (green).
4. Press the
LINEAR SINGLE assist key to set the linear single sweep mode.
The source shape indicator shows the staircase icon, and the status changes to
MOVE (blue).