EasyManuals Logo

Keysight Technologies B2900 Series User Manual

Keysight Technologies B2900 Series
234 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #83 background imageLoading...
Page #83 background image
Keysight B2900 User’s Guide, Edition 5 3-19
Installation
Connecting a DUT
Using the N1295A Test Fixture
Keysight N1295A is a test fixture which has four triaxial connectors and which
supports two 2-wire connections.
Requirements Keysight N1295A test fixture furnished with pin clip wire 4 ea. and pin plug
wire 2 ea.
Keysight N1294A-001 banana to triaxial adapter for 2-wire connections
Triaxial cable, 2 ea. for one 2-wire connection
Connections 1. Attach the banana to triaxial adapter to the B2900 source/measure terminals. See
Figure 3-6. If there is some space as “Bad connection”, the contact is not
enough.
2. Connect the triaxial cables between the adapter and N1295A as shown in Figure
3-7.
3. Connect the DUT between terminals 1 and 2 for the example in Figure 3-7.
Figure 3-6 Attaching Banana to Triaxial Adapter
Figure 3-7 Keysight N1295A Connection Example
NOTE Shielding
N1295A is equipped with a cover. To minimize the effect of ambient noise, close the
cover when performing measurements.
Bad connection
Good connection (no space)
1
1
2
3
4
3
2
4
± 42V 1.05 A Max
Low Force
High Force
DUT

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies B2900 Series and is the answer not in the manual?

Keysight Technologies B2900 Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelB2900 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals