6-14 Keysight B2900 User’s Guide, Edition 5
Function Details
Automatic Output-ON/OFF Function
Automatic Output-ON/OFF Function
The automatic output-ON/OFF function defines the source channel’s output
ON/OFF operation when the trigger system changes the status.
Automatic-ON function
If this function is enabled, the source/measure unit (SMU) automatically turns
the channel output ON just before the trigger system is initiated by an SCPI
command, not by a front panel operation.
Automatic-OFF function
If this function is enabled, the SMU automatically turns the channel output OFF
immediately when all trigger system changes the status from busy to idle.
To set the automatic output ON/OFF function, see “Output Connection dialog box”
on page 4-28.