Keysight B2900 User’s Guide, Edition 5 4-51
Front Panel Reference
Display key group
Display Preference dialog box
This dialog box is used to set the following parameters. The settings are saved to the
non-volatile memory.
V/I Limit for Each Polarity
Selects the Limit (Compliance) setup operation.
ON Individual setup ON
Positive and negative limits can be set individually.
OFF Individual setup OFF
± Limit value is set.
This function is available for the B2900 firmware revision 3.1 or later. And the
“SWS” license is required to enable this function. See “To Use Functions other
than Measurement” on page 5-5 to confirm the firmware revision and the license
applied to the B2900.
Also see “File key group” on page 4-43 for the license.
Immediate V/I Update by Knob
Selects the channel operation when the Source value or the Limit (Compliance)
value is changed by using the digit pointer and the rotary knob.
ON Real time update ON
Turning the rotary knob changes the display value and updates the
channel output in real time.
OFF Real time update OFF
Turning the rotary knob just changes the display value. The value is
applied to the channel output by pressing the rotary knob.
Restore Trigger Mode on Local
Stores the trigger mode setting on entering GP-IB remote state, ON or OFF.
ON Saves the trigger mode setting in the non-volatile memory on
entering GPIB remote state and restores the saved trigger mode
setting when returning to GPIB local mode.
OFF Disables this function.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies B2900 Series and is the answer not in the manual?

Keysight Technologies B2900 Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelB2900 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals