Keysight B2900 User’s Guide, Edition 5 4-53
Front Panel Reference
System key group
Auto CLR Displays the following softkeys for setting the automatic clear
of the timestamp. The present setting is indicated by an asterisk
on the softkey label.
OFF Disables the automatic clear of the timestamp.
ON Enables the automatic clear of the timestamp.
Start-up Displays the System Start-up dialog box which provides the following setup
parameters.
Power-on State Power-on state, RST, RCL0, RCL1, RCL2, RCL3, or RCL4
The power-on state can be selected from the factory default
reset condition (RST) and user conditions RCL0, RCL1,
RCL2, RCL3, and RCL4 which can be defined by using the
*SAV 0, *SAV 1, *SAV 2, *SAV 3, and *SAV 4 commands,
respectively, while the B2900 is in the remote mode.
Power-on Program Power-on program execution ON or OFF
If this parameter is ON, the power-on program is
automatically executed when the B2900 is turned on.
The power-on program can be defined by using the
:PROG:PON:COPY command while the B2900 is in the
remote mode.
Sound Displays the following softkeys to enable or disable the beep and sound. The present
setting is indicated by an asterisk on the softkey label.
OFF Disables the beep and sound.
ON Enables the beep and sound.
Language Displays the following softkeys for specifying the B2900 remote control command
set. The present setting is indicated by an asterisk on the softkey label.
Default Specifies the default command set which supports all B2900
functions.
2400 Specifies the conventional command set designed for using the
existing programs which you created to control existing
instruments, such as Series 2400 by Keithley Instruments, Inc.
Script Not effective.
The default command set and the conventional command set are available.