6-36 Keysight B2900 User’s Guide, Edition 5
Function Details
Initial Settings
Table 6-7 SMU Initial Settings
Program variables None Not changed
Program selection None
Program status Idle
Setup item Power on Reset
Low terminal state GROUNDED Not changed
Output state OFF
Output auto filter enable ON
Output filter enable ON
Output filter time constant 5 s
Output filter frequency 31.8309886 kHz
Output OFF state NORMAL
Output protection OFF
Source mode VOLTS
Source shape DC
Automatic output-on ON
Automatic output-off OFF
Voltage source 0 V
Voltage protection level 2 V
Voltage source auto ranging ON
Voltage source range 2 V
Voltage source range lower limit 0.2 V
Voltage source mode Constant voltage
Setup item Power on Reset