EasyManuals Logo

Keysight Truevolt Series User Manual

Keysight Truevolt Series
583 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #315 background imageLoading...
Page #315 background image
[SENSe:]CAPacitance:NULL:VALue:AUTO {ON|1|OFF|0}
[SENSe:]CAPacitance:NULL:VALue:AUTO?
Enables or disables automatic null value selection for capacitance measurements.
Parameter Typical Return
{ON|1|OFF|0}. Default: ON. 0(OFF)or1(ON)
Configure capacitance measurements, turn null function on, use the AUTO null function. Make two meas-
urements and send them to the instrument's output buffer:
CONF:CAP AUTO
CAP:NULL ON
CAP:NULL:VAL:AUTO ON
SAMP:COUN 2
READ?
Typical Response: +3.01963841E-10,+3.01566277E-10
l When automatic reference selection is ON, the first measurement made is used as the null value for all
subsequent measurements.[SENSe:]CAPacitance:NULL:VALue is set to this value. Automatic null
value selection will be disabled.
l When automatic null value selection is disabled (OFF), the null value is specified by:[SENSe:]CAPa-
citance:NULL:VALue.
l The instrument enables automatic null value selection when the null function is enabled
([SENSe:]CAPacitance:NULL:STATe ON).
l This parameter is set to its default value after a Factory Reset (*RST), Instrument Preset
(SYSTem:PRESet), or CONFigure function.
[SENSe:]CAPacitance:RANGe:AUTO {OFF|ON|ONCE}
[SENSe:]CAPacitance:RANGe:AUTO?
Disables or enables autoranging for capacitance measurements.
Parameter Typical Return
{OFF|ON|ONCE}. Default: ON. 0(OFF)or1(ON)
Configure capacitance measurements, turn autorange on, turn null function on, use the AUTO null function.
Make two measurements and send them to the instrument's output buffer:
CONF:CAP
CAP:RANG:AUTO ON
CAP:NULL ON
CAP:NULL:VAL:AUTO ON
SAMP:COUN 2
READ?
Typical Response: +3.01963841E-10,+3.01566277E-10
l Autoranging goes down a range at less than 10% of range and up a range at greater than 120% of
range. For capacitance measurements only, when autorange is off, the instrument does not report an
SCPI Programming Reference
316
Keysight Truevolt Series Operating and Service Guide

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Truevolt Series and is the answer not in the manual?

Keysight Truevolt Series Specifications

General IconGeneral
BrandKeysight
ModelTruevolt Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals