[SENSe:]CAPacitance Subsystem
This subsystem configures capacitance measurements.
Command Summary
[SENSe:]CAPacitance:NULL[:STATe]
[SENSe:]CAPacitance:NULL:VALue
[SENSe:]CAPacitance:NULL:VALue:AUTO
[SENSe:]CAPacitance:RANGe:AUTO
[SENSe:]CAPacitance:RANGe
[SENSe:]CAPacitance:SECondary
[SENSe:]CAPacitance:NULL[:STATe] {ON|1|OFF|0}
[SENSe:]CAPacitance:NULL[:STATe]?
Turns the null function on or off for capacitance measurements.
Parameter Typical Return
{ON|1|OFF|0}. Default: ON. 0(OFF)or1(ON)
Configure capacitance measurements, turn null function on, use the AUTO null function. Make two meas-
urements and send them to the instrument's output buffer:
CONF:CAP AUTO
CAP:NULL ON
CAP:NULL:VAL:AUTO ON
SAMP:COUN 2
READ?
Typical Response: +3.01963841E-10,+3.01566277E-10
l Enabling the scaling function also enables automatic null value selection ([SENSe:]CAPa-
citance:NULL:VALue:AUTO).
l To set a fixed null value, use: [SENSe:]CAPacitance:NULL:VALue.
l The instrument disables the null function after a Factory Reset (*RST), Instrument Preset
(SYSTem:PRESet), or CONFigure function.
SCPI Programming Reference
314
Keysight Truevolt Series Operating and Service Guide