EasyManua.ls Logo

Keysight Truevolt Series - Page 353

Keysight Truevolt Series
583 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
See Also
[SENSe:]{RESistance|FRESistance}:RANGe:AUTO
[SENSe:]{RESistance|FRESistance}:RANGe:AUTO {OFF|ON|ONCE}
[SENSe:]{RESistance|FRESistance}:RANGe:AUTO?
Disables or enables autoranging for all resistance measurements. Autoranging is convenient because it
automatically selects the range for each measurement based on the input signal. Specifying ONCE per-
forms an immediate autorange and then turns autoranging off.
This parameter is common to 2-wire and 4-wire resistance measurements. Using the FRESistance version
of this command or query is identical to using the RESistance version.
Parameter Typical Return
{ON|1|OFF|0}. Default: ON. 0(OFF)or1(ON)
Configure 2-wire resistance measurements and perform an immediate autorange. Make and read two meas-
urements:
CONF:RES
RES:RANG:AUTO ONCE
SAMP:COUN 2
READ?
Typical Response: +1.04530000E+03,+1.04570000E+03
l Autoranging goes down a range at less than 10% of range and up a range at greater than 120% of
range.
l With autoranging enabled, the instrument selects the range based on the input signal.
l
Selecting a fixed range ([SENSe:]<function>:RANGe) disables autoranging.
l This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset
(SYSTem:PRESet).
SCPI Programming Reference
354
Keysight Truevolt Series Operating and Service Guide

Table of Contents

Related product manuals