SYSTem:ACALibration:DATE?
Returns the date of the last autocalibration in the format yyyy,mm,dd.
Parameter Typical Return
(none) 2014,4,26
Return the autocalibration date:
SYST:ACAL:DATE?
l (34465A/70A only)
l You can perform this query regardless of whether the instrument is secured.
l The date is based on the instrument's real-time clock. Set the date for the instrument's real-time clock
with SYSTem:DATE.
l The real-time clock does not adjust itself for time zone changes or daylight savings time. Be sure to set
the date and time when you first receive your instrument.
l This setting is non-volatile; it will not be changed by power cycling or *RST or SYSTem:PRESet.
l Use *CAL? to perform autocal.
See Also
SYSTem:DATE
SYSTem:ACALibration:TEMPerature?
Returns the temperature of the last autocalibration in °C.
Parameter Typical Return
(none) +2.42850208E+001
Return the autocalibration temperature:
SYST:ACAL:TEMP?
l (34465A/70A only)
l You can perform this query regardless of whether the instrument is secured.
l This setting is non-volatile; it will not be changed by power cycling or *RST or SYSTem:PRESet.
l Use *CAL? to perform autocal.
Keysight Truevolt Series Operating and Service Guide
393