SYSTem:COMMunicate:LAN:TELNet:PROMpt "<string>"
SYSTem:COMMunicate:LAN:TELNet:PROMpt?
Sets the command prompt seen when communicating with the instrument via Telnet.
Parameter Typical Return
String of up to 15 characters
Default is Instrument Model Number: 34460A>, 34461A>, 34465A>, or
34470A>
"Command>"
Set the command prompt:
SYST:COMM:LAN:TELN:PROM "Command>"
l On the 34460A, this requires option 34460A-LAN or option 3446LANU.
l The instrument uses LAN port 5024 for SCPI Telnet sessions and port 5025 for SCPI Socket sessions.
l Telnet sessions are typically started from a host computer shell:
telnet <IP_address> <port>
For example:
telnet 169.254.4.10 5024
To exit a Telnet session, press <Ctrl-D>.
l This setting is non-volatile; it is not changed by power cycling, a Factory Reset (*RST), or an Instru-
ment Preset (SYSTem:PRESet).
l SYSTem:SECurity:IMMediate sets this parameter to its default value.
SCPI Programming Reference
414
Keysight Truevolt Series Operating and Service Guide

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Truevolt Series and is the answer not in the manual?

Keysight Truevolt Series Specifications

General IconGeneral
BrandKeysight
ModelTruevolt Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals