Medrad™ Intego™
PET Infusion System
56
16. Route the Saline Tube along the SAS Track, starting at the SAS Confluence Holder and
moving towards the Saline Pump. Be sure to press the tubing firmly into the SAS Track.
17. Press the Saline Tube into the Saline Tube Holder between the Saline Pump and SAS
Track.
INT SYS 200
INT SYS 100
WARNING: Improper injection dosage or radiation spillage may result if the
Shielded Chamber Lid is closed onto the tubing set. Ensure that the tubing set
and all administration set components are properly installed before closing the
Shielded Chamber Lid.
18. Once the SAS has been installed, ensure all components are laying flat prior to closing
the lid.