4 - Daily Setup
a. Center the Saline Tube on the Saline Pump rollers. The Saline Pump will not pump
fluid properly if the tubing falls below the rollers.
CAUTION: Incorrect installation of the SAS may result in improper functioning
of the Intego™ PET Infusion System.
b. Ensure the Saline Disconnect Luer rests against the edge of the Saline Pump.
CAUTION: Incorrect installation of the SAS may result in improper functioning
of the Intego™ PET Infusion System.
c. Center the Saline Tube in the Saline Pump door to prevent the tubing from being
pinched when the door is closed.
CAUTION: Incorrect installation of the SAS may result in improper functioning
of the Intego™ PET Infusion System.
WARNING: Air embolism or improper injection dosage may occur if Saline Line
is damaged. Make certain the Saline Tube is not pinched within the Saline Pump as
this may damage the administration set.