EasyManuals Logo

MICRO-EPSILON CS02 User Manual

MICRO-EPSILON CS02
116 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #16 background imageLoading...
Page #16 background image
Page 16
Functional Principle, Technical Data
capaNCDT 6200
Demodulator module DL62xx
The demodulator module DL62xx consists of an internal preamplifier, demodulator, output stage and A/D
converter. The internal preamplifier generates the distance dependent measuring signal and amplifies it. De-
modulator and output stage convert the measuring signal in a standardized voltage and current signal. The
measuring values can be processed digitally with the help of the A/D converter.
The trim potentiometer zero allows a special zero adjustment of the analog output signals, see Fig. 3.
i
Output voltage can achieve up to 15 VDC, if the sensor is disconnected respectively exceedance of
measuring range.