Leak and Compliance Tests Preoperative Tests
4 - 12 046-004667-00 A7™ Operating Instructions
Time out
Example:
Test results cannot be shown due to an internal
communication error.
Select Retry to repeat the Automatic Circuit Leak Test &
Compliance test.
or
Select Override to skip the test.
Backup Flow Control Enabled
Example
For the A7 System, the backup flow control is active.
Follow on-screen instructions to troubleshoot the
problem.
or
Select Accept to proceed to operational mode using
previous compliance values.
RESULTS COMMENTS/OPTIONS
TABLE 4-3 Automatic Circuit Leak and Compliance Test Results