A7™ Operating Instructions 046-004667-00 9 - 25
Product Specifications Response Time of the A7 to an Increase in O2 Concentration (From 21% to 90%/100%)
9.18 Response Time of the A7 to an Increase in O
2
Concentration (From 21% to 90%/100%)
Ventilator setting: TV=500 ml, Freq=10 bpm, I:E=1:2
Fresh Gas(L/min) O
2
Flush Response Time
From 21% to 90%
Response Time
From 21% to 100%
1.0 Without O
2
Flush 15 minutes 50 minutes
With O
2
Flush 1 minute 2 minutes
5.0 Without O
2
Flush 2 minutes 6 minutes
With O
2
Flush 1 minute 1 minute
10.0 Without O
2
Flush 1 minute 2 minutes
With O
2
Flush 1 minute 1 minute
TABLE 9-38
Response Time of the A7 to an Increase in O
2
Concentration (From 21% to 90%/
100%)