A7™ Operating Instructions 046-004667-00 C - 13
Parameters and Factory Defaults Ventilation Parameter Relationships
C.10 Ventilation Parameter Relationships
VENTILATION
MODE Parameter
Parameter
Relationship Equation (s)
VCV Rate
4 ≤ Rate ≤ 100
Vt
20 ≤ Vt ≤ 1500
Plimit Plimit ≥ PEEP+5
10≤ Plimit ≤ 100
SIMV-VC Rate
4 ≤ Rate ≤ 100
Vt
20 ≤ Vt ≤ 1500
ΔP ΔP ≤ Plimit-PEEP
3 ≤
ΔP ≤ 50
Plimit Plimit ≥ PEEP+5
Plimit ≥
ΔP+PEEP
10 ≤ Plimit ≤ 100