Linked Ventilation Parameter Parameters and Factory Defaults
C - 12 046-004667-00 A7™ Operating Instructions
SIMV-PC Pinsp PLAT
or 80% PEAK
or last value
** *
Rate ****
Tinsp ****
PS ***PS = On
ΔP ****
Trigger ****
PEEP ****
Tslope ****
PS Min Rate ****
ΔP ****
Trigger ****
Peep ****
Tslope ****
Apnea Ti ****
CURRENT
VENTILATION MODE &
PARAMETERS
AFFECTED
PREVIOUS VENTILATION MODE
VCV SIMV-VC PCV SIMV-PC PS
* The parameter value is shared between the previous and current ventilation modes.