A7™ Operating Instructions 046-004667-00 C - 11
Parameters and Factory Defaults Linked Ventilation Parameter
C.9 Linked Ventilation Parameter
The table below lists how parameter values are affected when changing ventilation modes. For
example, ventilation modes that share the same parameters may also share the same parameter
values when changing from one ventilation mode to the other. Other parameters may have their
values set differently when changing ventilation modes.
CURRENT
VENTILATION MODE &
PARAMETERS
AFFECTED
PREVIOUS VENTILATION MODE
VCV SIMV-VC PCV SIMV-PC PS
VCV Vt —* Measured Vt
or last value
**
Rate —* * **
I:E —* * **
Tpause —* * **
PEEP —* * **
Plimit —* * **
SIMV-VC Vt *—Measured Vt
or last value
**
Rate *—***
Tinsp *—***
Tpause *—***
PEEP *—***
Plimit *—***
PS *—**PS = On
ΔP *—***
Trigger *—***
Tslope *—***
PCV VtG ****
Pinsp PLAT
or 80% PEAK
or last value
*—**
Rate ****
I:E ****
PEEP ****
PlimVG If VtG=OFF,
then Pinsp.
If VtG is a
value, then
last value of
PlimVG.
If VtG=OFF,
then Pinsp.
If VtG is a
value, then
last value of
PlimVG.
—If VtG=OFF,
then Pinsp.
If VtG is a
value, then
last value
of PlimVG.
If VtG=OFF,
then Pinsp.
If VtG is a
value, then
last value of
PlimVG.
Tslope ****
* The parameter value is shared between the previous and current ventilation modes.

Table of Contents

Other manuals for Mindray A7

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray A7 and is the answer not in the manual?

Mindray A7 Specifications

General IconGeneral
TypeAnesthesia Workstation
BrandMindray
ModelA7
Display15-inch color touchscreen
Oxygen FlushYes
Gas InletsO2, N2O, Air
Ventilation ModesSIMV, PSV
Breathing SystemIntegrated
Gas MonitoringO2, CO2, N2O
Monitoring ParametersFiO2
CO2 MonitoringMainstream or Sidestream
Power Supply100-240 V, 50/60 Hz

Related product manuals