Silencing Alarms Alarms and Messages
6 - 8 046-004667-00 A7™ Operating Instructions
6.4 Silencing Alarms
When an alarm condition occurs and the alarm audio sounds, the user can select the Silence softkey
at the bottom of screen to silence the alarm audio. In silenced status, all the alarm indicators work
normally except audible alarm tones.
When the Silence softkey is selected, all active alarms are silenced and the icon on the left side of the
alarm message changes to indicate that the alarm is silenced. When the 120 second silence icon
appears, the audio alarms are silenced for 120 seconds, after which the audio alarms resume.
If you select the Silence softkey while all alarms are silenced, then the audio alarms will resume
immediately.
NOTE: A new alarm will sound if that alarm occurs while the system is in a
silenced state. If this occurs, you can select the Silence softkey again to
silence the new alarm and reset the silence countdown timer to 120
seconds.
FIGURE 6-5 Alarm audio-paused
Alarm audio-paused Icon
120 second countdown time icon

Table of Contents

Other manuals for Mindray A7

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray A7 and is the answer not in the manual?

Mindray A7 Specifications

General IconGeneral
TypeAnesthesia Workstation
BrandMindray
ModelA7
Display15-inch color touchscreen
Oxygen FlushYes
Gas InletsO2, N2O, Air
Ventilation ModesSIMV, PSV
Breathing SystemIntegrated
Gas MonitoringO2, CO2, N2O
Monitoring ParametersFiO2
CO2 MonitoringMainstream or Sidestream
Power Supply100-240 V, 50/60 Hz

Related product manuals