Cleaning and Disinfection Maintenance
7 - 18 046-004667-00 A7™ Operating Instructions
7.9.8 Bag Arm
1. At the base of the bag arm, locate the retaining ring. Turn the ring counterclockwise until it
is no longer threaded. Lift the bag arm from the breathing system block (see FIGURE 7-15).
FIGURE 7-15 Bag Arm Removal
2. Cleaning
a. Clean the bag arm with a soft, lint-free cloth and a recommended cleaning agent (see
TABLE 7-2 on page 7-8). Allow it to dry thoroughly.
b. If disinfecting the bag arm, continue with step 3, otherwise skip to step 4.
3. Disinfection
NOTE: Ensure that the bag arm has been cleaned as described in step 2 before
disinfecting.
Use an approved disinfecting agent (see TABLE 7-3 on page 7-9 and TABLE 7-4 on page 7-9) for
the bag arm while adhering to facility policies and procedures.
4. Reassemble the bag arm to the breathing system. Prior to use after cleaning or disinfecting,
power up the system and follow the on-screen prompts to perform the leak test and the
compliance test (see section 4.5 (page 4-8) "Leak and Compliance Tests").
Retaining
Ring
Bag Arm

Table of Contents

Other manuals for Mindray A7

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray A7 and is the answer not in the manual?

Mindray A7 Specifications

General IconGeneral
BrandMindray
ModelA7
CategoryMedical Equipment
LanguageEnglish

Related product manuals