Alarm and Prompt Messages Alarms and Messages
6 - 20 046-004667-00 A7™ Operating Instructions
6.6.2.2 CPU Board Runtime Alarm
Keyboard Self
Test Error
Keyboard self-test Error High Startup only Non-
Functional
Keyboard
Keyboard Self
Test : Time out
Keyboard self-test
result cannot be
obtained due to
communication error.
High Startup only Non-
Functional
Keyboard
External AG Self
Test Error
If the module sends the
ErrorMsg, except for
data limit error and
unspecified accuracy,
"External AG Self Test
Error" will be triggered.
Low Startup only All AG Module
Internal AG
Error O
2
If the module sends the
ErrorMsg, except for
data limit error and
unspecified accuracy,
"Internal AG Error O
2
"
will be triggered.
Low Startup only All AG Module
External AG:
Time out
External AG selftest
result cannot be
obtained due to
communication error.
Low Startup only All AG Module
Internal AG:
Time out
Internal AG selftest
result cannot be
obtained due to
communication error.
Low Startup only All AG Module
MESSAGE CAUSE
ALARM
PRIORITY
MACHINE
MODE
WHEN
CHECKED
STARTUP
RESULT IF
FAIL REMARK
TABLE 6-4 Startup Alarm Messages
MESSAGE CAUSE
ALARM
PRIORITY
MACHINE
MODE WHEN
CHECKED
DISABLE IN
STANDBY
MODE
IP Address Conflict The IP address of the
machine is the same as
the IP address of another
device in the local
network.
Medium Runtime No
Fan Failure Speed of the fan ≤ 20% of
normal speed
Medium Runtime No
Fan Failure O
2
Speed of Module Rack
fan < 3640
Medium Runtime No
TABLE 6-5 CPU Board Runtime Alarm Messages