Setup Menu Parameters and Factory Defaults
C - 4 046-004667-00 A7™ Operating Instructions
C.3 Setup Menu
PARAMETER RANGE DEFAULT
General Tab: Breathing System Warmer On, Warmer Off Warmer On
General Tab: Gas Bench Flow
Rate
Adult watertrap:
Low (120 ml/min),
Med (150ml/min),
High (200 ml/min)
Infant watertrap:
Low (70 ml/min),
Med (90 ml/min),
High (120 ml/min)
Low (120 ml/min)
General Tab: Balance Gas AIR, N
2
O, None AIR
General Tab: Fresh Gas Control Total Flow, Direct Flow Total Flow
Display Tab: Pressure Display Mean, PLAT PLAT
Display Tab: Plimit Line On/Off On
Display Tab: Screen Brightness level 1-10 5
Display Tab: Key Click Volume level 1-10 3
Display Tab: CO
2
Placement Top, Bottom Top
Display Tab: CO
2
Scale 0-40 mmHg,
0-60 mmHg,
0-80 mmHg
0-60 mmHg
Display Tab: Waveform Display Volume, Flow Flow
System Tab: Language CHINESE,ENGLISH, FRENCH, SPANISH,
PORTUGUESE, RUSSIAN, TURKISH,
DUTCH
ENGLISH
System Tab: Default Settings:
Default Patient Size
Adult, Pediatric, Infant Infant
System Tab: Default Settings:
Default Vent Mode
VCV, SIMV-VC, PCV, SIMV-PC, PS VCV
System Tab: Manage Defaults Save as O.R. Defaults, Load O.R.
Defaults, Restore Partial Defaults,
Import Defaults,
Export Defaults
Save as O.R. Defaults
System Tab: Change Password — 1234
System Tab: Units: Pressure cmH
2
O, hPa, mbar cmH
2
O
System Tab: Units: CO
2
mmHg, kPa, % mmHg
System Tab: Clear History On, Off Off
System Tab: Optimizer:
Optimizer
On, Off On
System Tab: Time Settings:
Daylight Savings
Manual, Auto Manual
System Tab: Network: This
Machine: Configure Ethernet: IP
Address
0 - 255 192.168.23.250